Urine Sediment Analysis System

BW-3000 Auto Urine Sediment Analyzer

← Back to Products
Product Image
Overview

BW-3000 Auto Urine Sediment Analyzer

● Test Principle

By using flow counting technology, medical image scanning technology and intelligent vision recognition technology, urine sediment will be 3D automatically located, tracked, scanned, identified and classified.

 

● The highest 4-channel design in the industry

4-channel injection design, channel settling, cleaning and other detection functions fully guarantees the speed and efficiency of specimen detection.

 

● Sampling System

The sample waiting area can accommodate 60 specimens at a time. The specimen is scanned and positioned automatically, without manual guarding and without centrifugation.

 

● Flushing System

The use of high-precision, PTFE-coated suction needles makes the instrument pipeline cleaning more thoroughly, and eliminates cross-contamination as much as possible.

 

● Auto Focus Function

Automatic focus correction of the microscope ensures clear and complete sample shooting, which reduces maintenance costs.

 

● Self-learning and Training Function

The target of misidentification of the instrument can be modified by the operator. After training, the memory can be stored and automatically modeled.

Specifications
  • ● Test Principle: Flow cell digital imaging technology, medical image fusion technology, artificial intelligence identification technique.

    ● Channel: 4 fixed channels in flow counting cell.

    ● Sample Volume: Minimum volume: 3 ml; aspiration volume: 1 ml.

    ● Test Speed: 120 samples/hour.

    ● Aspiration Volume: 1 ml.

    ● Test Sensitivity: 5 cells/ul.

    ● Accuracy: ≥ 95%.

    ● Repeatability: CV < 10%.

    ● Cross Contamination Rate: < 0.05%.

    ● Data Storage: 200,000.

    ● Report Model: International general quantitative unit: unit/ul & HPF, complete and illustrated report, with urine chemistry test results and urine sediment test results.

    ● Environment: Temperature 15°C–35°C, optimum temperature 20°C–25°C, relative humidity ≤ 75%.

    ● Power Supply: AC 110/230V, 50–60Hz.

    ● Dimensions (Main Unit): 660mm×440mm×570mm (L × W × H).

    ● Weight: 48 kg.