Urine Sediment Analysis System

BW-3000 Auto Urine Sediment Analyzer

Back to Urine Sediment Analysis System
Product Image
Overview

BW-3000 Auto Urine Sediment Analyzer

Uses flow counting chamber, medical image scanning, and AI vision recognition to automatically locate, track, scan, identify, and count urine formed elements in 3D.

● Product Advantages

10-position automatic track sample rack with continuous loading; sample area holds up to 60 samples.

Fully automatic operation, no manual supervision needed.

No centrifugation; direct raw urine testing with automatic aspiration, mixing, and drainage.

PTFE-coated high-precision sampling needle for thorough cleaning and low cross-contamination.

4-channel sampling design enables simultaneous sedimentation, cleaning, and testing for high efficiency.

Misidentified targets can be corrected manually; system learns and builds models after training.

● Application

Hospital labs, outpatient/emergency departments, physical exam centers.Primary healthcare and third-party labs.

Specifications

● Detection Principle

Four-channel quantitative flow counting chamber, planar flow technology, medical image information fusion technology, and AI machine vision recognition technology.

● Test Items

Normal RBC, Standing RBC, Dysmorphic RBC, Normal WBC, WBCC, Pus Cells, Hyaline Casts, Granular Casts, WBC Casts, Squamous Epithelial Cells, Normal Epithelial Cells, Sperm, Calcium Crystals, Oxalate Crystals, Bacteria, Yeast, Amorphous Urates, Mucus, etc.

● Test Speed

120–180 tests/hour

● Data Storage Capacity

200,000 records

● Environment

-Temperature: 15°C–35°C

-Optimal temperature: 20°C–25°C

-Relative humidity: ≤75%

● Other Parameters

-Sample volume: minimum 3 ml

-Aspiration volume: 1 ml

-Sensitivity: <10 /μl

-Recognition rate: >95%

-Repeatability (CV): <10%

-Carryover rate: <0.05%

-False negative rate: <3%

-Power supply: AC220V ±22V, 50Hz ±5Hz

-Dimensions: 660mm × 440mm × 570mm